This instrument is suitable for measuring the thermal conductivity of refractory insulation materials, ceramic fibers, felts, textiles, boards, bricks, and other materials at various temperatures. With an optional surrounding frame, it can also be used to measure powdered materials. It is widely employed in material analysis and testing by universities, research institutions, quality‑inspection agencies, and manufacturing plants.
This instrument conforms to the following standards: ASTM C518-04, Standard Test Method for Steady-State Thermal Flux and Thermal Transmission Properties Using the Heat Flow Meter; ASTM C182-88 (2004), Standard Test Method for Thermal Conductivity of Insulating Refractory Bricks; ASTM C202-93 (2004), Standard Test Method for Thermal Conductivity of Refractory Bricks; ASTM C417-05, Standard Test Method for Thermal Conductivity of Amorphous Refractories; ASTM C201-93 (2004), Standard Test Method for Thermal Conductivity of Refractory Materials; GB 10295-88, Determination of Steady-State Thermal Resistance and Related Properties of Thermal Insulation Materials (Heat Flow Meter Method); YB/T 4130-2005, Test Method for Thermal Conductivity of Refractory Materials (Water‑Flow Plate Method); and GB/T 17911-2018, Test Method for Thermal Conductivity of Refractory Ceramic Fiber Products, among others.
Technical Specifications
| Thermal conductivity range | 0.0010–3.0 W/m·K, displayed to four decimal places |
| Test accuracy | Better than 5% |
| Hot surface temperature range | 100-1000℃ |
| Calorimetric temperature control method | Semiconductor cooling; optional constant-temperature water bath cooling available at an additional cost. |
| Heat-preserving and temperature-control method | Semiconductor cooling |
| Calorimeter plate dimensions | 100×100 (mm) |
| Heat flux meter dimensions | 100×100 (mm) |
| Number of temperature measurement points on the specimen | 4 o'clock |
| Specimen atmosphere environment | Air |
| Pressure range | 0–1000 N, with electric pressurization via a geared motor and precise measurement using a high-accuracy pressure sensor. |
| Thickness measurement range | 0–100 mm, measured using a high-precision displacement sensor. |
| Thickness Correction | Instantaneous correction of the displacement sensor |
| Specimen dimensions | 230×230 to 250×250 (mm); thickness 5 or 10 to 100 (mm) |
| Measurement control method | The system performs fully automated testing and data processing, and can generate test reports for printed output. |
| Host power supply | Voltage: 220 V; Frequency: 50 Hz; Total Power: ≤ 5 kW |
| Cooling method | Cooling is achieved using tap water; a thermostatic water bath can be optionally configured upon customer request. The thermostatic water bath offers temperature control within the range of 0–99.99°C, with an additional charge. |
| Instrument Configuration | One test host, one set of test software, one set of communication interface and data cable, and one roll of rubber hose; the buyer shall provide its own computer and printer. |
Related Product
HC-40 Concrete Multi-functional Strength Tester
The HC-40 concrete multi-functional strength tester is a high-performance testing device integrating multiple advanced technologies, specifically designed for evaluating the strength and quality of concrete.
FAQs
Customer Reviews
European Plastic Pipe Manufacturer|Purchase: Ring Stiffness & Hydrostatic Burst Tester
2026.06.05
★★★★★
Middle East Construction Lab|Purchase: Universal Electronic Testing Machine & Geotextile Test Equipment
2026.06.05
★★★★★
Southeast Asia Rubber & Polymer Factory|Purchase: Vicat Softening Point & Tensile Testing Machine
2026.06.05
★★★★★
Consult
We will contact you within one working day. Please pay attention to your email.